Medical Electric Equipment Safety Performance Integrated Testing Apparatus (DAZ-Y)
Гарантия возврата денег, если товар не подойдёт
Описание
The apparatus designed and manufactured according with GB9706.1-1995, whose technical capacities ∈dexes totally accord with GB9706.1-1995, IEC601-1: 1998. It can test earthing impedance, power, leakage current, dielectric intensity, remainder voltage, and test automatically in the setting order by PLC( earthing impedance- leakage current -power-remainder voltage-dielectric intensity). LCD's menus is controlled by touch key, and measuring data can be printed by the printer in the apparatus. Technical data: 1. Earthing impedance test: Output current 1A, 10A, 25A, error: ± 3%± 2Testing Range: 1~250mΩ, error: ± 3%± 2Alarm figure: 0.05Ω, 0.1Ω, 0.15Ω, 0.2Ω, 0.5Ω, error: ± 3%± 22. Leakage current test: Testing Range: 0.5μA~10mA, error: ± 5%, MD impedance, DC-1MHz Virtual Value. Alarm figure: 0.01mA, 0.05mA, 0.1mA, 0.5mA, 1mA, 2.5mA, 5mA, 10mA, error± 5%. Voltage of power supply: AC 121V, 242V error: ± 2%, Output Power: 2KW (if 242V)Voltage of adding press: AC 121V, 242V, 250V error: ± 2%, Output Power: 100W(250V)3. Power test: Single-phase virtual power: 50~2000W(if 242V), error: ± 3%± 2Output voltage: 99V, 110V, 121V, 198V, 220V, 242V, error: ± 3%± 2 4. Leakage current test: Voltage of power supply: AC 121V, 242V error: ± 3%± 2Testing Range: 3~60V error ± 1VGive an alarm if output impedance >100MΩ, >60V 5. Dielectric intensity test: Output voltage: 500V~5KV, error: ± 3%± 2, become the set figure in 10s Alarm figure: 2mA, 5mA, 10mA, 15mA, error: ± 3%± 2
Свойства товара
- Model NO: DAZ-Y
- Trademar: HY
- Specificatio: IEC601-1:1988